รายละเอียดเพิ่มเติม
Lepu Neo ECG S120
ECG S120 AI-ECG Platform is an artificial intelligence (AI) electrocardiogram-ECG-assisted analysis and diagnosis system independently developed by Lepu Medical.
Product Description
- High Accuracy Rate
Test by 1,000,000 independent MIT test data, the average accuracy rate of the AI-ECG platform reach 95.2%.
- High Analysis Speed
Take 1s for automatically resting ECG analysis. The time saved can reduce the overall time of clinical ECG analysis.
- Comprehensive Diagnosis
Support 16 types of cardiac classification, 104 types of ECG diagnostic classifications.
Product Features
• 7″ high-resolution color touch screen, easy to operate. Portable design, compact in size.
• Can be powered by an external DC power supply, a built-in rechargeable lithium battery.
• Support synchronous acquisition and display of the 9/12-lead waveform, as well as heart rate detection.
• Support automatic pacing detection and marking.
• Support auto, RR analysis, HRV, medicine test, ECG event mode.
• Provide 4 sampling modes: pre-sampling, real-time sampling, periodic sampling and trigger sampling.
• Support automatic pacing detection.
• Input patient information via a virtual alphanumeric keyboard and barcode scanning.
• Freeze the ECG waveform on the screen.
• Output files in multiple formats, such as Carewell ECG, PDF, BMP, HL7, DICOM, SCP.
• Store, preview, review, edit, export, upload, print and search patient data.
• Support wireless transmission of ECG data via WiFi and mobile networks.
• Support laser printer.
• Export patient data to USB flash disk via a USB connector.
• Support the user login permission control, use password or account & password authentication to use the device.
• Support online and offline login to the device, and view the historical patient data of the department according to the login account.
• Support connection with AI-ECG PLATFORM in achieving intelligent diagnosis (Optional).
Technical Specifications
Neo ECG S120 | ||
Main unit | HR range | 30bpm~300bpm |
Accuracy | ±1 | |
Leads | 9,12 lead synchronous acquisition | |
A / D conversion | 24 bits | |
Sampling rate | 32000 samples/Sec | |
Common mode rejection ratio | ≥140dB (AC filter on)≥120dB (AC filter off) | |
Time constant | ≥5s | |
Frequency response | 0.01HZ~350HZ (+0.4db-3.0db) | |
Sensitivity | Auto, 2.5mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV, 40 mm/mV, less than ±5% error | |
Fliter | AC filter: 50Hz, 60Hz, Off | |
EMG filter: 25Hz, 35Hz, 45Hz, Off | ||
ADS filter: 0.01 Hz, 0.05 Hz, 0.32 Hz, 0.67 Hz | ||
Low pass filter: 75Hz,100Hz,150Hz,300Hz,Off | ||
Paper speed | 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s and 50mm/s, less than ±3% error | |
Input Impedance | ≥100MΩ (10Hz) | |
Input Circuit Current | ≤10nA | |
Calibration voltage | 1mV±2% | |
Depolarization voltage | ±900mV, ±5% | |
Noise | ≤12.5μV | |
Amplitude quantisation | 0.95 μV/LSB | |
Recovery time after defibrillation discharge | <10s | |
Pacer pulse display | Pacing pulse with amplitude of ±2mV~±700mV, duration of 0.1ms~2.0ms, A-5 rise time of less than 100µs, and frequency of 100/min can be displayed on the ECG recording. | |
Minimum detectable signal | 20μVp-p | |
Size & Weight | 7” Tablet | Size: 197mm(L) * 112.4mm(W) * 26.1mm(H) Net Weight: 0.8Kg |
Analysis algorithm | Glasgow |