Logo Loading

Autoclave Sturdy SA-260MA

 

รายละเอียดเพิ่มเติม

เครื่องนึ่งฆ่าเชื้อเครื่องนึ่งฆ่าเชื้อด้วยไอน้ำชนิดแนวนอน (Autoclave) Sturdy SA-260MA

Sturdy’s advanced tabletop autoclaves for solid, wrapped, tubal and liquid (optional) sterilization. These autoclaves use Microprocessor design, providing smart logic control for precision temperature monitoring and a full range of settings for diagnostics, drying and preprogrammed sterilization presets. The most main feature is a powerful vacuum pump for thorough inside-out sterilization of tubal instruments.

คุณลักษณะ Sturdy SA-260MA :

Chamber Capacity (L) 24
Maximum Instrument Length (mm) 350
Maximum Load (unpouched) (g) 7,000
Maximum Load (pouched) (g) 1,300
External Dimensions (mm) 655 (D) x 533 (W) x 442 (H)
Chamber Size (mm) Ø 260 mm x 450 mm (D)
Gross Weight (kg) 54
Voltage/Wattage (Heater) 220V-240V AC, 50/60Hz 1800W
Flash program: 2300W
Fuses 15A x 2
Flash program: 20A × 2
Water Reservoir Capacity (ml) 4200
Water Capacity per Cycle (ml) 280 ~ 350
Sterilization Temperature (°C) 121 / 134
Working Environment Indoor use;
Under 1,000m (altitude);
Temperature 5°C to 40°C;
Relative Humidity 80%RH@31°C to Relative Humidity 50%RH@40°C;
Voltage fluctuation ±10 %;
Transient overvoltages category II;
Pollution degree 2
Transportation Conditions -10°C to 70°C, 10%RH to 90%RH
Storage Conditions -10°C to 50°C, 10%RH – 70%RH
Over Pressure Protection Safety valve 2.55 bar (2.6 kgf/cm²)
Over Pressure Indication Pressure protection switch with warning LED
Over Temperature Indication Independent temperature protection device with warning LED
Water Level Indication Water level sensor in reservoir and chamber with warning LED
Door Lock Indication Micro switch sensor with warning LED
Pressure Display Analog pressure gauge
Function Display LED
Recorder
Program Selections Wrapped/unwrapped, Dry/no dry, Re-Dry (or Liquid, Flash **)*.
* Optional function by order request
** The Max. 1.5 kg loading (unwrapped, solid load) for Flash program. (The performance of Flash program may be varied by operation environment)
Temperature Selection 121°C/134°C
Drying Function 0, 15, 30, 35, 40 min. (Re-Dry: 10 min.)
Additional controls PRION, B.D. Test, Pump Test, Reset/Stop, Vacuum release, Emergency