รายละเอียดเพิ่มเติม
เครื่องจี้ไฟฟ้า Surtron Touch 200
เครื่องจี้ไฟฟ้า Surtron Touch 200 is an High Frequency electrosurgical equipment which is suitable for precision monopolar and bipolar surgery and micro-surgery without tissue alterations.
SURTRON® TOUCH can supply HF operative powers either for programmable time as short as hundreds of a second.
The unit through its performances allows pure CUT, cut-coagulation BLEND, incision with reduced production of eschar ENHANCED, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG, bipolar CUT and COAG. The digital reading of the delivered power and the overseeing through microcontroller of the operational functions, assure the absolute reliability of the conditions of work.
SURTRON® TOUCH allows a highly professional surgery thanks to the user-friendly and safety solutions normally used.
The connection of neutral electrode is constantly monitored. Safety control of patient/plate contact using split neutral electrode. The possibility to control by the handle the output functions as well as the delivery of output power, allows to implement the surgical operation without turning away the surgeon attention from the surgical field.
UNDERWATER CUTTING AND COAGULATION
SURTRON TOUCH 200
FEATURES
Max power PURE CUT | 200 W – 300 Ω |
Max power BLEND CUT | 150 W – 300 Ω |
Max power ENHANCED CUT | 150 W – 300 Ω |
Max power BIPOLAR TUR | 200 W – 50 Ω |
Max power BIPOLAR PURE | 120 W – 50 Ω |
Max power FORCED COAG | 150 W – 200 Ω |
Max power SOFT COAG | 100 W – 200 Ω |
Max power FULGURATION COAG | 100 W – 1000 Ω |
Potenza max BIPOLAR COAG | 120 W – 50 Ω |
Max power BIPOLAR SEALING | 200 W – 50 Ω |
Working frequency | 360 kHz |
RF power pulse emission time | 10ms-30sec |
RF power pulse interval | 1-30 |
Patient circuit | F |
Selectable input voltage | 90-240 Vac |
Mains Frequency | 50-60 Hz |
Electrical input power | 750 VA |
Size WxHxD | 370x145x300 mm |
Weight | 5.8kg |
CONTROLS
Patient/Plate circuit monitoring
Output power monitoring
Self check control
SAFETY
EN60601-1
EN60601-1-2
EN60601-2-2
Electrical classification: I CF
MDD 93/42 / EC: II b